| Model Number |
401763 |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Valvular Insufficiency/ Regurgitation (4449)
|
| Date of Event |
02/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a transcatheter aortic valve replacement procedure when the valve was deployed, a premature ventricular contraction occurred.During the procedure, the temporary pacing catheter failed to capture, causing the valve to deploy approximately 2-3 mm distal to the intended site.A transthoracic echocardiogram and cineangiography were performed which revealed a perivalvular leak.It was felt that failure of the temporary pacing catheter to maintain capture led to the complication / inappropriate placement of valve.Fda ref: mw5115460.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported valvular insufficiency remains unknown.
|
| |
|