• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CORAIL AMT SZ10 135 HIGH COL; CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CORAIL AMT SZ10 135 HIGH COL; CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM Back to Search Results
Model Number L971110
Medical Device Problem Codes Use of Device Problem (1670); Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 02/27/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Patient's painful hip was revised due to loose stem.No sign of infection.Surgeon believed stem was potted distally and that led to loosening.Implants are unavailable for return.No surgical delay was reported.Doi: unknown.Dor: (b)(6) 2023.Affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary - no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot - the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL AMT SZ10 135 HIGH COL
Common Device Name
CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16592057
Report Number1818910-2023-06396
Device Sequence Number3021686
Product Code MEH
UDI-Device Identifier10603295455295
UDI-Public10603295455295
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K192946
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberL971110
Device Catalogue NumberL971110
Device Lot Number9810986
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2023
Supplement Date Received by Manufacturer03/23/2023
Initial Report FDA Received Date03/22/2023
Supplement Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-