MEDTRONIC MED REL MEDTRONIC PUERTO RICO EVERA¿ XT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number DDBB1D1
Device Problems Failure to Charge (1085); Delayed Charge Time (2586)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2023
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was inappropriately shocked by the right ventricular (rv) lead due to oversensing.The rv lead triggered a lead integrity alert (lia) for high rate non-sustained (hr-ns) and sensing integrity counter (sic) episodes.The lead alerted for oversensing of noise and high/undefined bipolar pacing impedance.The lead had high pacing thresholds.The lead had a confirmed fracture on the pace/sense portion of the lead.The pace/sense was capped leaving the high voltage coils in use.A new pace/sense lead was implanted.The lead remains in use.It was further reported that the right atrial (ra) lead has undersensing.The lead remains in use.It was further reported that the implantable cardioverter defibrillator (icd) had a charge circuit timeout.The charge time was longer than expected.The device was removed and a new device was implanted.No further patient complications have been reported, as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERA¿ XT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16593102
MDR Text Key311788492
Report Number3004209178-2023-03876
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00643169018099
UDI-Public00643169018099
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/28/2015
Device Model NumberDDBB1D1
Device Catalogue NumberDDBB1D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694958 LEAD, 407645 LEAD
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexMale
Patient Weight124 KG
-
-