| Model Number |
C4120 |
| Medical Device Problem Codes |
Material Separation (1562); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Procedure performed: lap hernia repair event description: the lattice pad popped off one of the jaws during surgery.The pad was retrieved from the patient.The case was completed with a new device.No patient injury and the patient is doing fine.Product available for return.Intervention: used a new one to complete the case patient status: fine.
|
| |
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation without a latis pad.Visual inspection confirmed the complainant¿s experience of pad detachment.Based on the condition of the returned unit, the pad detached due to adhesive failure, likely resulting from usage during the procedure.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|
Event or Problem Description
|
|
Procedure performed: lap hernia repair event description: the lattice pad popped off one of the jaws during surgery.The pad was retrieved from the patient.The case was completed with a new device.No patient injury and the patient is doing fine.Product available for return.Intervention: used a new one to complete the case patient status: fine.
|
| |
|
Search Alerts/Recalls
|