MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES C4120, 38CM GRASPER REPOS CART, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES C4120, 38CM GRASPER REPOS CART, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC Back to Search Results
Model Number C4120
Medical Device Problem Codes Material Separation (1562); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
Procedure performed: lap hernia repair event description: the lattice pad popped off one of the jaws during surgery.The pad was retrieved from the patient.The case was completed with a new device.No patient injury and the patient is doing fine.Product available for return.Intervention: used a new one to complete the case patient status: fine.
 
Additional Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation without a latis pad.Visual inspection confirmed the complainant¿s experience of pad detachment.Based on the condition of the returned unit, the pad detached due to adhesive failure, likely resulting from usage during the procedure.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Procedure performed: lap hernia repair event description: the lattice pad popped off one of the jaws during surgery.The pad was retrieved from the patient.The case was completed with a new device.No patient injury and the patient is doing fine.Product available for return.Intervention: used a new one to complete the case patient status: fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C4120, 38CM GRASPER REPOS CART, 10/BX
Common Device Name
LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key16593165
Report Number2027111-2023-00391
Device Sequence Number5088880
Product Code NWV
UDI-Device Identifier00607915116866
UDI-Public(01)00607915116866(17)260105(30)01(10)1477125
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2026
Device Model NumberC4120
Device Catalogue Number101472471
Device Lot Number1477125
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2023
Supplement Date Received by Manufacturer03/16/2023
Initial Report FDA Received Date03/22/2023
Supplement Report FDA Received Date07/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-