| Model Number |
VICM512.6 |
| Medical Device Problem Codes |
Misfocusing (1401); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Disturbances (2140)
|
| Date of Event |
01/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm vicm512.6 implantable collamer lens of a diopter of -9.0 was implanted into the patient's left eye (os) on (b)(6) 2022.On (b)(6) 2023, the lens was removed and replaced for a different model due to refractive surprise and blurred vison.Cause of the event is unknown.Reportedly, "surgical incision induced more astimatism than expected," and "refractive miss - patient had 1.D of astigmastism after spherical icl placement, despite having only 0.5d pre-op so i exchanged the spherical icl for a toric." the status of the eye is reported as, "heathy.".
|
| |
|
Additional Manufacturer Narrative
|
|
Weight-race:unk.Work order search found no similar complaints.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d9- returned to manufacturer: "13-mar-2023" should be corrected to "10-mar-2023." b5- "cause is unknown and the device failed to perform as intended." device evaluation: h3 - type of investigation code: 10- device evaluation: the lens was returned in a vial with liquid.Visual inspection found no visible damage.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h3- device evaluation: dimensional and functional inspection found the lens to be within specifications.Corrected data: b5- the reporter indicated that the surgeon implanted a 12.6mm vicm5_12.6 implantable collamer lens of diopter -9.00d into the patients left eye (os) on (b)(6) 2022.The patient experienced refractive surprise and blurred vision.Reportedly "1d astigmatism after spherical icl placement, despite having only 0.50d pre-op".On (b)(6) 2023 the lens was exchanged for the different model/ same length and the problem was resolved.Status of the eye is noted as "healthy".The cause for this event was unknown and the device failed to perform as intended.- "surgical incision induced more astigmatism than expected".H6- medical device problem code: "2017" should be removed.Claim# (b)(4).
|
| |
|