| Model Number |
VICM5_13.2 |
| Medical Device Problem Codes |
Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that an implantable collamer lens was implanted into the patient's eye.The reporter notes a planned lens explant and replacement due to excessive vault.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5-the reporter indicated that a 13.2mm vicm5_13.2 implantable collamer lens of -11.50 diopter was implanted into the patients left eye (os) on (b)(6) 2023.On (b)(6) 2023 the lens was explanted due to excessive vault and elevated iop (intraocular pressure).A replacement lens of shorter length was later implanted and the problem resolved.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h6: 1494-off label- progressive myopia d9: return date: 05-apr-2023 (b)(4).
|
| |
|