Model Number MMT-105NNBLNA |
Device Problems
Material Too Rigid or Stiff (1544); Obstruction of Flow (2423); Wireless Communication Problem (3283)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 02/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Information received by medtronic indicated that the customer reported hyperglycemia with a blood glucose value of 309 mg/dl.Troubleshooting was performed and found that the inpen was not dispensing the insulin.The customer was not reporting any symptoms related to the blood glucose value.The customer was treated with a manual injection.The insulin exited after changing the new cartridge and repeating the prime 5 times.The customer also reported the inpen was not communicating with the mobile app and the customer reported the dose dial knob was difficult to turn and there was an additional resistance experienced when turning dose knob/dial greater than 4 units.No harm requiring medical intervention was reported.The customer will discontinue using the inpen.The inpen will be returned for analysis.
|
|
Manufacturer Narrative
|
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Medical device problem code has been updated and provide with this report in section h6.
|
|
Manufacturer Narrative
|
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.The inpen paired to the commercial app.Inpen passed baseline and wireless functionality.App logbook displayed: 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u.Inpen passed front cap investigation.Pending further investigation performed in san diego location.In conclusion: per san diego analysis: inpen passed base line functionality test and displacement dose accuracy.Missing injection foot.No other problems found with this inpen.Therefore, the customer concern of inpen not pairing and difficult to dial/dose could not be confirmed.However, injection foot damaged was confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|