• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INMODE LTD BODYTITE; INMODE RF SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INMODE LTD BODYTITE; INMODE RF SYSTEM Back to Search Results
Model Number AG606969A
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Full thickness (Third Degree) Burn (2696)
Date of Event 07/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Left posterior upper arm burned while using bodytite technology.
 
Additional Manufacturer Narrative
Third-degree burn (full thickness) located in the postero-medial aspect of the left arm, approximately 3 x 2 cm in size.The event occurred in (b)(6) 2022 and is only now being reported.Following the use of the bodytite in various positions and in both supine and prone decubitus.The patient was subsequently treated with 5 sessions of laser over the scar area.The last documented follow-up session dates back to (b)(6) 2022 where the issue has not been resolved.Although at present the patient would in all likelihood have recovered and largely healed; however, not having the photos of the current state and in consideration of the entity of the event (third degree burn) it was decided to report the case to the competent authorities.
 
Event or Problem Description
Left posterior upper arm burn while using bodytite technology.
 
Additional Manufacturer Narrative
Follow up 07/02/2023.1.Investigation conclusions : user error excluding technical failure and being the setting of parameters in accordance with the protocol, it is a question of a wrong maneuvering, approaching too close to the port of access and lingering too long while the energy was being supplied.Summary of related investigations: excess of energy in the surface area due to incorrect movement technique with the cannula.2.Medical device problem code - changed to 2017.3.Type of investigation - changed to 10.4.Investigation findings - changed to 4248.5.Investigation conclusions - changed to 61.6.Added section g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BODYTITE
Common Device Name
INMODE RF SYSTEM
Manufacturer (Section D)
INMODE LTD
tavor building (pob 533)
shaar yokneam
yokneam northern, north 20692 0
IS  206920
Manufacturer (Section G)
INMODE LTD
tavor building (pob 533)
shaar yokneam
yokneam northern, north 20692 06
IS   2069206
Manufacturer Contact
eran krieger
tavor building (pob 533)
shaar yokneam
yokneam northern, north 20692-06
IS   2069206
MDR Report Key16597458
Report Number3010511300-2023-00207
Device Sequence Number14582337
Product Code GEI
UDI-Device Identifier07290016633337
UDI-Public07290016633337
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K182325
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Remedial Action Inspection,Notification,Patient Monitoring
Type of Report Initial,Followup
Report Date (Section B) 07/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAG606969A
Device Catalogue NumberAG606969A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer06/14/2023
Initial Report FDA Received Date03/23/2023
Supplement Report FDA Received Date07/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age44 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-