| Model Number |
SA60AT |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Eye Injury (1845)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A non-health care professional reported during the intraocular lens implantation they were unable to successfully load the lens into the cartridge.The lens was final loaded using the long angle forceps.Then the injector probe kept passing over or under the iol without advancing it.The injector plunger was slightly bent up.The lens was inspected and there were no scratches on the lens.So they enlarged the corneal incision to about 6mm, and manually delivered the lens into the anterior chamber.Then closed the incision with 3 interrupted 10-0 nylon sutures.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis; the lens remains implanted.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|