This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: investigation type codes were added: 3331, 4109 and 4111.H6: investigation findings code was added: 3252.H6: investigation conclusions codes were added: 4307.H10: narrative/data was updated.Zimvie received one (1) oss410, (osseotite implant 4 x 10mm) for evaluation.Visual evaluation was performed, a fracture has been identified on the implants threads.Dhr review was completed for the subject lot number 2019080040.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2019080040 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation were material selection of implant inadequate (unable to withstand occlusal forces or long term loading) therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the threads.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|