OSCOR INC. GUIDESTAR¿ STEERABLE GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number D141103
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conclusion not yet available, evaluation in process.A follow-up will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Event or Problem Description
Unusually, the sheath drew a lot of air.After a number of aspirations, we decided to change the sheath because the risk for the patient was too high.There was a 3 minute delay in the procedure to obtain a new sheath which resolved the issue.
 
Additional Manufacturer Narrative
The following sections were updated in follow-up 1.One 14f guidestar steerable guiding sheath was received without the dilator.There were no other accessories.Traces of blood were found on and inside the sheath.According to the event description summary, the sheath drew a lot of air.The hemostatic valve was viewed under magnification and looked normal and intact with no anomalies.The sheath was aspirated with the tip submerged into water and a syringe connected to the stopcock.When the plunger of the syringe was pulled back, water was drawn into the sheath fully and smoothly.No bubbles were seen inside the sideport tubing.The sheath was manually leak tested with the syringe connected to the stopcock and the tip occluded.The sheath did not exhibit leakage when tested using this method.The sheath was then leak tested using the iso 11070 for liquid leakage test method.The device was occluded at the distal tip and pressurized to 38kpa and held at this pressure for 30 seconds and no leakage was observed from the hemostatic valve.The sheath was further pressurized beyond 300kpa and no leakage was observed from the stopcock, sideport tubing or through the handle.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Complaint review of lot # dp-16255 identified no additional report for similar event.Returned device analysis revealed the sheath was within manufacturing specifications.The sheath aspirated fully and smoothly, and no bubbles were observed inside the sideport tubing.The hemostatic valve looked normal and was intact.The sheath did not leak when tested manually and also met the iso leakage test method requirement.According to the dhr, the sheath passed all in-process and final inspection steps including visual, mechanical, dimensional and leak testing.No manufacturing defects were found.Per qa procedure (destino steerable guiding sheath in-process and final inspection) sample size: ansi z 1.4, gen level i, normal, aql 1.5 normal perform leak test according to procedures based on available leak tester.The leak test is performed by manufacturing personnel at 100% and is observed by quality assurance personnel at aql level ansi z 1.4, gen level i, normal, aql 1.5 normal.Per ifu; the steerable sheath must be thoroughly flushed with either saline or heparinized saline and free of air prior to use to avoid air embolism to the patient.Aspiration and flushing of the sheath should be performed frequently to help minimize the potential for air embolism.For injecting or aspirating through the sheath, use the sideport only with stopcock.Prior to infusion, remove all air using the sideport.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDESTAR¿ STEERABLE GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683
Manufacturer Contact
daniel naut
3816 desoto blvd.
palm harbor, FL 34683
7279372511
MDR Report Key16599568
Report Number1035166-2023-00029
Device Sequence Number6093703
Product Code DYB
UDI-Device Identifier00885672010911
UDI-Public00885672010911
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140406
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2025
Device Model NumberD141103
Device Catalogue NumberD141103
Device Lot NumberDP-16255
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2023
Supplement Date Received by Manufacturer08/09/2023
Initial Report FDA Received Date03/23/2023
Supplement Report FDA Received Date08/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-