COVIDIEN 760 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number G-760220DIUA-EN
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a patient information and other relevant history relevant history cannot be provided due to china's personal information protection law.Section e: initial reporter was not provided.Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the medtronic service engineer.This event was reported from the china food and drug administration (cfda), adverse (b)(4) rather than reported directly to medtronic by the customer.The device was not available for evaluation.It was reported that the customer repaired the unit by themselves by replacing the 'power module' and returned the unit to normal use.No product was returned for evaluation or made available to medtronic.With the information available the likely cause of the issue was isolated to a potential faulty in the 'power module'.However, inve stigation also determines the unit reached the end of service life (eos).The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality s pecifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported by the china food and drug administration (cfda), adverse reaction event center of shanghai, and not reported directly to medtronic by the customer, that while in use on a patient, the main screen of this 760 ventilator went black and the unit automatically stopped running.The patient was manually ventilated and transferred to an alternate ventilator without injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
760 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16599756
Report Number8020893-2023-00155
Device Sequence Number12593944
Product Code CBK
UDI-Device Identifier10884521180932
UDI-Public10884521180932
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K984379
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG-760220DIUA-EN
Device Catalogue NumberG-760220DIUA-EN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2023
Initial Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/11/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1