• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP RADIOLUCENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP RADIOLUCENT Back to Search Results
Model Number 3034-00
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 02/23/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The slight deviation (<10%) of the pressure reading cannot be the cause of the slippage.Since the surface of the product has different types of damages (scratches and impact / pressure marks) in several points, we assume that the general condition of the product is conditioed by improper handling by the customer.It is also suspected that the crack in the fixed arm at the interface with the extension arm was caused by external impact(s).In general, it cannot be excluded that further of the detected deviations, such as that of the clamping screw, were caused by improper handling of the device and/or by external impacts of force.We recommend using our skull pins in combination with our skull clamps.Risks due to not permitted system combinations with third party components could not be excluded, as in this case.
 
Event or Problem Description
Customer informed our service department on the 24th of february that one of our products was involved in a case where a laceration occured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP RADIOLUCENT
Common Device Name
DORO® SKULL CLAMP RADIOLUCENT
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
MDR Report Key16599805
Report Number3003923579-2023-00009
Device Sequence Number14993686
Product Code HBL
UDI-Device Identifier04250435502181
UDI-Public04250435502181
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 02/24/2023
Report Date (Section F) 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3034-00
Device Catalogue Number3034-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2023
Distributor Facility Aware Date02/24/2023
Device Age3 YR
Event Location Hospital
Date Report to Manufacturer02/24/2023
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/23/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-