|
Initial reporter address: (b)(6).Qc was acceptable on the date of the event.Five sample-related alarms were observed on the alarm trace data.Precision testing was performed.The original sample was stored for 3 days at 2 8 °c.Product labeling states the sample is: "stable for 5 hours at 20 25 °c, 12 hours at 2 8 °c, 3 months at 20 °c (± 5 °c).Freeze only once.Ck mb stability is extremely temperature-dependent.A ck mb decrease of > 10 % can occur after the sample has stood for 1 hour at 32 °c." the patient suffers from myeloproliferative neoplasms (mpn), and when using heparin tubes the blood viscosity increases.This may lead to a drop at the tip of the sample probe, which could lead to a high result.Clots could also influence the results.Based on the data provided, the cause of the event could not be determined.A general reagent issue can be excluded.Qc data does not suggest a general reagent or instrument issue.The investigation did not identify a product problem.
|