CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE) Back to Search Results
Model Number VPR-GW-FLEX14
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/23/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The guide wire was returned to csi.An extra spring tip was observed on the core shaft.Visual examination did not observe any fracture or damage on the guide wire.The extra spring tip was due to an outside supplier component issue.The reported event of guide wire fracture was not confirmed.The material inspection report for this guidewire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Event or Problem Description
Following treatment of a severely calcified lesion, angiographic imaging revealed the tip of the viperwire advance guide wire fractured in the ankle in the posterior tibial artery.A non-csi support catheter was used to retrieve the fractured tip.A shepherd guide wire was used to perform balloon angioplasty.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE)
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key16600902
Report Number3004742232-2023-00076
Device Sequence Number6193891
Product Code MCW
UDI-Device Identifier10852528005718
UDI-Public(01)10852528005718(17)241130(10)463246-1
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberVPR-GW-FLEX14
Device Catalogue Number7-10041-03
Device Lot Number463246-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Initial Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-