|
SYNTHES GMBH PROTECTION SLEEVE FOR FNS INSRT INSTS; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
|
Back to Search Results |
|
| Model Number |
03.168.008 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device photos.Visual analysis of the photo revealed that the protection sleeve for fns insrt insts had the distal tip heavily deformed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the protection sleeve for fns insrt insts.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, a couple of fns items were found damaged after deconning.During manufacturer's investigation of the returned device it was identified that the protection sleeve for fns insrt insts had the distal tip heavily deformed.This report is for one (1) femoral neck system insertion handle this is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|