|
The device was returned to stryker sustainability solutions for evaluation.Functional testing was performed via inline testing.Given the reported event, per re-inspection documentation, the device was found to be eligible for rejection based on the relevant reject code of cp, curve out of plane.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to an electrical failure as a result of mishandling subsequent to distribution from stryker.The instructions for use (ifu) state: - ep catheters should be used only by or under the supervision of an appropriately trained physician using proper procedures and techniques.- do not autoclave catheter.- do not use for electrical ablation.- if electrocardiogram equipment is used for placement of the catheter, the equipment must be front-end isolated of have an isolated patient cable.Current leakage from the electrocardiogram monitor must not exceed 10 microamps.- avoid excessive torque, stretching, kinking and/or bending of catheter, as this may interfere with distal tip shaping or cause damage to internal electrode wires.- avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.Handle catheter with care to avoid improper electrical functioning.- standard electrical grounding precautions should be observed when using electrical recording or stimulation equipment.- avoid excessive contact of handpiece with fluids, as this could adversely affect the electrical performance of the catheter.The reported event will continue to be monitored through post-market surveillance.
|