| Model Number |
FIBERTAG TIGHTROPE WITH FLIPCUTTER III |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 03/06/2023, it was reported by a sales representative via phone that an ar-1288rtt-fc3 fibertag tightrope with flipcutter iii broke into pieces.This occurred during an acl reconstruction on (b)(6) 2023 when going to flip-cut the tunnels the flip cutter hinge broke off while reaming.All the fragments that were produced were retrieved.The case was completed using an ar-1204ff flipcutter.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to misaligned insertion and/or prying/leveraging during use.
|
| |
|