BOSTON SCIENTIFIC CORPORATION EMBOLD FIBERED DETACHABLE COIL SYSTEM; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83910
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation by manufacturer: the product was returned to boston scientific for analysis.The delivery wire and the nitinol shaft were returned.The returned delivery wire showed that it was not broken at the laser cut perforations.The delivery wire was returned with the proximal and distal coupler attached.The coupler showed no damage.The coil was not connected to the distal coupler.The coil was delivered into the patient.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Event or Problem Description
It was reported that the delivery wire broke, and a device fragment was left inside the patient.An 8x20 embold fibered detachable coil system was selected for use an embolization procedure.The target location was a pseudoaneurysm in the pancreaticoduodenal.The vessel was mildly tortuous.During the procedure, the physician retracted the coil due to a sizing preference.While trying to retract the coil into the sheath, there was resistance, and the polymer shaft of the delivery wire broke.The physician was able to use saline to deploy the coil with the distal fragment of the delivery wire still attached to the coil.The coil and fragment of the delivery wire were left inside the target location, and no attempts were made to retrieve the device fragment.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOLD FIBERED DETACHABLE COIL SYSTEM
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16610101
Report Number2124215-2023-13815
Device Sequence Number5131731
Product Code KRD
UDI-Device Identifier08714729983620
UDI-Public08714729983620
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/12/2025
Device Model Number83910
Device Catalogue Number83910
Device Lot Number0029579225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/27/2023
Initial Report FDA Received Date03/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-