| Model Number |
DBP-200SOLID145 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The oad was returned with the guide wire engaged.Significant resistance was noted when attempting to remove the guide wire.Visual analysis confirmed fractured driveshaft filars at the proximal side of the crown, however, the driveshaft was not fully separation.Scanning electron microscope analysis of the fractured filars revealed evidence of fatigue which is an indication that the driveshaft was spun in a high stress environment.Diagnostic analysis identified a stall event.The exact root cause of the reported events were unable to be determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
|
| |
|
Event or Problem Description
|
|
A diamondback 360 peripheral orbital atherectomy device (oad) was used to treat the common femoral artery.The device was retracted to the lesion in the common iliac artery.Two treatments on low speed were performed.When changing the speed to medium, the oad stopped spinning.The oad was difficult to remove and seemed to be stuck on the viperwire advance guide wire.The oad and guide wire were removed.Damage was observed on the driveshaft near the crown.A new guide wire was used and the procedure was complete with percutaneous transluminal angioplasty.The patient was stable.
|
| |
|