• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT Back to Search Results
Model Number 990063-020
Medical Device Problem Codes Signal Artifact/Noise (1036); Compatibility Problem (2960); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the 990063-020 mapping catheter with lot number 225151080 was returned and analyzed.Visual inspection of the loop segment area showed the loop was intact with no apparent issues.No damage was observed along with the tip/loop section of the mapping catheter.Visual inspection of the pebax segment area showed pebax tubing was kinked/twisted at approximately 2.71 inches from loop distal tip.The user may have experienced insertion difficulties due to this issue when introducing the mapping catheter into the balloon catheter.Visual inspection of the electrodes showed the electrodes were intact with no apparent issues.All electrodes exist on the loop section and no cosmetic issue or anomalies were identified.Visual inspection of the introducer showed the introducer was intact with no apparent issues.No damage or any other issue was observed along with the introducer.Pin to pin continuity test failed due to an open circuit at electrodes # 4,#5 ,#6,#7,#8.Due to this, the user may have experienced interference issues.Dissection of the suspected electrode related to the noise revealed the electrode wires were broken from the welding at electrodes 4, 5, 6 ,7, 8.In conclusion, the mapping catheter failed the returned product inspection due to the tip/loop kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, there was resistance felt when the mapping catheter was brought back into the sheath.The signal was then lost.The mapping catheter was adjusted without resolve.The mapping catheter was moved around the heart and signal was further lost.The mapping catheter was replaced to resolve the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER - 20 MM
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16613498
Report Number9617601-2023-00055
Device Sequence Number17426128
Product Code DRF
UDI-Device Identifier00643169467385
UDI-Public00643169467385
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102588
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2024
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number225151080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Initial Report FDA Received Date03/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
-
-