• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 20 CM, INLINE GRIP; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 20 CM, INLINE GRIP; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0520IC
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation of tissue pad peeling was not confirmed.In addition, the probe broke 15 millimeters from the tip of the probe.There were scratches around the broken part.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
An olympus employee reported on behalf of a customer, the thunderbeat tissue pad came off during surgery.The unspecified therapeutic procedure was completed using a replacement device.There was no reported patient harm associated with this event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1) during the output activation in seal & cut mode, the probe was contacting hard tissue, metal objects or surgical instruments.2) scratches were generated on the probe.3) a force to activate the output in seal & cut mode, or a force to grasp the tissue was applied to the probe.Therefore, cracks were generated at the scratched area.4) a force was applied to the probe causing it to break.The following information is included in the instructions for use (ifu): ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, other instruments, or forceps, and others.Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/split/protruding/partial separating of the tissue pad.In turn, the probe tip may break before displaying an error window or generating an alarm tone.¿ ¿do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.¿ olympus will continue to monitor the performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 20 CM, INLINE GRIP
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16616280
Report Number9614641-2023-00425
Device Sequence Number20047369
Product Code GEI
UDI-Device Identifier04953170308710
UDI-Public04953170308710
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132703
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0520IC
Device Lot Number27K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2023
Supplement Date Received by Manufacturer04/06/2023
Initial Report FDA Received Date03/24/2023
Supplement Report FDA Received Date04/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-