• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH QUICK COUPLING FOR DRILL BITS F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH QUICK COUPLING FOR DRILL BITS F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.750
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported from india that during an unspecified surgical procedure, it was discovered that the front portion of the quick coupling device came out.It was not reported if there were any delays in the surgical procedure or if a spare device was available.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The initial reporter's phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: this device was returned for evaluation.A visual and functional assessment was performed which determined that the device passed all pre-repair diagnostic assessments and no failures were identified.Therefore, the reported condition was not confirmed, and an assignable root cause was not determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICK COUPLING FOR DRILL BITS F/BPL
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key16617288
Report Number8030965-2023-03661
Device Sequence Number6149429
Product Code GEY
UDI-Device Identifier07611819137196
UDI-Public07611819137196
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number530.750
Device Lot Number8119786
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2023
Supplement Date Received by Manufacturer03/27/2023
05/09/2023
Initial Report FDA Received Date03/27/2023
Supplement Report FDA Received Date04/04/2023
05/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/14/2012
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-