| Model Number |
200-72-09 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Joint effusion (4849)
|
| Date of Event |
03/30/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Additional information received by the legal department on (b)(6) 2023 :
on (b)(6) 2023 , the patient underwent a left total knee replacement.On (b)(6) 2017, the patient underwent revision of the exactech knee device secondary to polyethylene liner wear with resulting osteolysis, liner breakage, and aseptic loosening.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Procodes: prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer.Concomitant medical products: 2429019 200-02-32 - three peg patella 32mm; 2559397 204-04-22 - trapezoid tibial tray sz 1f/2t, 2f/2t; 2382142 230-02-02 - optetrak asy,cr cemented femoral, sz 2.Initial mdr and fu1 reported under mfr# 1038671-2017-00216 reported under historical system, prior to electronic complaint handling system.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported in experience was likely the result of excessive overall posterior tibial slope of the tibial insert (as defined by the combination of the posterior slope of the proximal tibial cut and the built-in slope of the insert) and/or axial rotation misalignment between the femoral component and the tibial tray, which allowed for excessive posterior contact between the femoral component and the tibial insert mainly on the medial aspect leading to catastrophic wear of the uhmwpe tibial insert.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|