• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CR TIBIAL INSERT SZ 2, 9MM, SLOPE ++; SEE H10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CR TIBIAL INSERT SZ 2, 9MM, SLOPE ++; SEE H10 Back to Search Results
Model Number 200-72-09
Medical Device Problem Codes Material Integrity Problem (2978); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Joint effusion (4849)
Date of Event 03/30/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Additional information received by the legal department on (b)(6) 2023 : on (b)(6) 2023 , the patient underwent a left total knee replacement.On (b)(6) 2017, the patient underwent revision of the exactech knee device secondary to polyethylene liner wear with resulting osteolysis, liner breakage, and aseptic loosening.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
Procodes: prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer.Concomitant medical products: 2429019 200-02-32 - three peg patella 32mm; 2559397 204-04-22 - trapezoid tibial tray sz 1f/2t, 2f/2t; 2382142 230-02-02 - optetrak asy,cr cemented femoral, sz 2.Initial mdr and fu1 reported under mfr# 1038671-2017-00216 reported under historical system, prior to electronic complaint handling system.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported in experience was likely the result of excessive overall posterior tibial slope of the tibial insert (as defined by the combination of the posterior slope of the proximal tibial cut and the built-in slope of the insert) and/or axial rotation misalignment between the femoral component and the tibial tray, which allowed for excessive posterior contact between the femoral component and the tibial insert mainly on the medial aspect leading to catastrophic wear of the uhmwpe tibial insert.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CR TIBIAL INSERT SZ 2, 9MM, SLOPE ++
Common Device Name
SEE H10
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32563
Manufacturer Contact
kate jacobson
MDR Report Key16617385
Report Number1038671-2023-00512
Device Sequence Number7325082
Product Code JWH
UDI-Device Identifier10885862041210
UDI-Public10885862041210
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K082022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2018
Device Model Number200-72-09
Device Catalogue Number200-72-09
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2023
Supplement Date Received by Manufacturer01/14/2025
Initial Report FDA Received Date03/27/2023
Supplement Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/05/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-