| Model Number |
200-63-09 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Metal Related Pathology (4530); Swelling/ Edema (4577)
|
| Date of Event |
07/25/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Additional information received by the legal department on 28 feb 2023: on (b)(6) 2012, the patient underwent a right total knee replacement with an exactech knee device.On (b)(6) 2017, approximately 5 years later the patient underwent revision of the right exactech knee device secondary to significant polyethylene liner wear with resulting metallosis, osteolysis, and synovitis.It is also stated that despite undergoing the revision surgery, the patient experiences daily knee pain and discomfort which limit activities of daily living and recreation and impacts quality of life.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Procodes: prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer.Concomitant medical products: 2350164 200-02-35 - three peg patella 35mm; 2199224 204-04-33 - trapezoid tibial tray sz 3f/3t; 1304943 230-03-03 - optetrak asy,cr cemented femoral, sz 3.Follow up type: initial and fu1 fda mdr reported under mfr# 1038671-2017-00535, this is being filed, with additional information and this is from a historical case before electronic complaint handling system.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|