| Model Number |
SUTURE ANCHOR, BIO-CORKSCREW FT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee surgery when inserting the anchor the screw broke.All fragments were retrieved.According to the surgeon, no harm to the patient, operator, or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Upon visual investigation, it was noted that the implant was broken.The broken pieces were not returned for investigation.The method used to prepare the bone and the bone quality encountered during the procedure was not provided.The most likely cause can be attributed to improper bone preparation or prying/leveraging the screw during insertion.Functional testing was not performed due to the damage to the device.
|
| |
|