Model Number 19191
Device Problems Loss of or Failure to Bond (1068); Unintended Movement (3026)
Patient Problems Hemorrhage/Bleeding (1888); Hyperglycemia (1905)
Event Date 03/25/2023
Event Type  malfunction  
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.It was reported that the cannula had dislodged from the infusion site.This condition could interrupt insulin delivery and contribute to hyperglycemia.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Event Description
It was reported that the patient's blood glucose level was greater than 250 mg/dl.The pod was worn between 5 and 24 hours on the hip/buttocks.It was noted that the adhesive was not secure and the cannula dislodged from the site.Hyperglycemia treated with a new pod.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16619923
MDR Text Key312048620
Report Number3004464228-2023-08171
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385081120033
UDI-Public(01)20385081120033(11)220317(17)230917(10)L73107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/17/2023
Device Model Number19191
Device Catalogue NumberZXP425
Device Lot NumberL73107
Was Device Available for Evaluation? No
Date Manufacturer Received03/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/17/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-