| Model Number |
IPN028442 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reported issue: blood leakage was detected through the red/blue connection of the tubing.The system was changed and the issue was resolved.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The device history record of batch number 74k2200575 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).One unit of catalog number a-6000-08lf (pe adult-ped dry/ wet lf 6/cs) with lot number 74j2201063 was returned and sent to the manufacturing site for analysis.The manufacturing site reports the red connector from the ats connector was not received.A device history record review was performed and no relevant findings were identified.The customer complaint cannot be confirmed since the defective sample was received incomplete.The red connector form ats connector was not returned.
|
| |
|
Event or Problem Description
|
|
Reported issue: blood leakage was detected through the red/blue connection of the tubing.The system was changed and the issue was resolved.No injury reported.
|
| |
|
Search Alerts/Recalls
|