• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF 6/CS; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028442
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
Reported issue: blood leakage was detected through the red/blue connection of the tubing.The system was changed and the issue was resolved.No injury reported.
 
Additional Manufacturer Narrative
Qn# (b)(4).The device history record of batch number 74k2200575 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Additional Manufacturer Narrative
(b)(4).One unit of catalog number a-6000-08lf (pe adult-ped dry/ wet lf 6/cs) with lot number 74j2201063 was returned and sent to the manufacturing site for analysis.The manufacturing site reports the red connector from the ats connector was not received.A device history record review was performed and no relevant findings were identified.The customer complaint cannot be confirmed since the defective sample was received incomplete.The red connector form ats connector was not returned.
 
Event or Problem Description
Reported issue: blood leakage was detected through the red/blue connection of the tubing.The system was changed and the issue was resolved.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED DRY/ WET LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194332672
MDR Report Key16622377
Report Number3004365956-2023-00023
Device Sequence Number7324784
Product Code KDQ
UDI-Device Identifier34026704631231
UDI-Public34026704631231
Combination Product (Y/N)N
Initial Reporter CountrySP
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2025
Device Model NumberIPN028442
Device Catalogue NumberA-6000-08LF
Device Lot Number74K2200575
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2023
Supplement Date Received by Manufacturer05/01/2023
Initial Report FDA Received Date03/27/2023
Supplement Report FDA Received Date06/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-