The device was returned and the complaint was confirmed.Examination of the returned device was completed by nuvasive and the fracture pattern identified is consistent with excessive off-angled force.Examination and review of all information received suggests excessive off-angled force was applied by user and considered the root cause of the event.The unretrieved tip in the patient is not a concern for migration as it was reportedly cold-welded into the screw head.No additional investigation required.Manufacturer review: review of the device history record notes no material non-conformance¿s, no manufacturing errors, nor discrepancies with respect to material type, treatments, dimensions that may have caused or contributed to this mode of failure.Parts met acceptance criteria upon release.Labeling review: "warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." "do not implant the instruments: complications to the patient may include, but are not limited to: breakage of the device, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration.Breakage could cause injury to the patient." "pre-operative warnings: the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.Care should be used during surgical procedures to prevent damage to the devices and injury to the patient." "intra-operative warnings: the physician should take precautions against putting undue stress on the spinal area with instruments.Any surgical technique should be carefully followed.It is important that the surgeon exercise extreme caution when working in close proximity to vital organs, nerves, or vessels, and that the force applied to the instrumentation is not excessive, to prevent potential injury to the patient.Over-bending, notching, striking, and/or scratching of implants with any instrument should be avoided to reduce the risk of breakage.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
|