| Model Number |
24705 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.Returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.The device was inflated to rated burst pressure and then deflated without any issues.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon deflation issue occurred.The patient presented with lower limb peripheral artery disease (pad).The 70% stenosed target lesion was located in the mildly tortuous and none calcified dorsalis pedis artery.A 2.0mm x 150mm x 150cm coyote balloon catheter was advanced for dilatation.However, the balloon would not deflate properly.The balloon was slightly deflated and was able to be removed intact.The procedure was completed successfully.There were no patient complications nor injuries reported.
|
| |
|