MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-5-15-3D-SS
Medical Device Problem Codes Detachment of Device or Device Component (2907); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that a axium coil prematurely detached in the microcather.During procedure the coil prematurely detached into microcatheter's distal portion before complete the wall apposition, then healthcare provider (hcp) carefully removed full system, like guide catheter, microcatheter with damaged coil from patient's right middle cerbral artery (mca) 2 portion.There was friction or difficulty during delivery.The hcp repositioned the coil 2 times.The hcp did not attempt to detach the coil.The hcp did rotate the delivery pusher during the procedure.Continuous flush was administered during the procedure.The devices were prepared according to the instructions for use (ifu).The patient was being treated for an unruptured, saccular aneurysm in the right mca 2 region.The max diameter was 7.1mm and the neck diameter was 3,3mm.Patient blood flow was high and vessel tortuosity was severe.No patient symptoms or complications were reported.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that the catheter resistance was in the distal portion.The coil come out of the introducer sheath.After removal coil was little kinked.The catheter was used by competition company.2/3 time attempts were made by using the instant detacher method.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key16623051
Report Number2029214-2023-00543
Device Sequence Number5577463
Product Code KRD
UDI-Device Identifier00847536032040
UDI-Public00847536032040
Combination Product (Y/N)N
Initial Reporter CountryBG
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2024
Device Model NumberAPB-5-15-3D-SS
Device Catalogue NumberAPB-5-15-3D-SS
Device Lot NumberB191559
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2023
Supplement Date Received by Manufacturer05/02/2023
Initial Report FDA Received Date03/27/2023
Supplement Report FDA Received Date05/04/2023
Date Device Manufactured04/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient Weight81 KG
-
-