Investigation found the power amplifier was damaged. for the report of overheating: thermal imaging was used to verify the outside temperature of the device while operating, outside temperature while charging, and the internal temperature while operating.Results are as follows: outside thermal temperature while operating: 35.4°c.Outside thermal temperature while charging: 34.5°c.Internal thermal temperature while operating: 47.5°c.The outside thermal temperature while charging was 34.5°c, which is below the product requirements specification of 41°c (106°f).Based on the thermal imaging, there is no evidence of thermal damage and the product met temperature specifications.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in blown power amplifier.Blown power amplifier rates remain acceptably low; thus, capa is not required.Blown power amplifier rates will continue to be tracked and trended.
|