| Model Number |
ES89162400 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to available information, this device required replacement due to a tubing break.The right cylinder tubing was ruptured.The reservoir was retained and reused.No other adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the available information the titan was revised due to ruptured right cylinder tubing.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot confirm any observations and cannot comment on the condition of the device.If the device becomes available, or additional information is received, quality will re-evaluate this complaint in accordance to procedures.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Additional Manufacturer Narrative
|
|
A titan pump and two cylinders were returned for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.A separation, surrounded by abrasion, was noted on the shorter exhaust tube of the pump.This is a site of leakage.No functional abnormalities were noted with either cylinder.Based on examination of the returned product, it was concluded that the abrasion marks noted on both exhaust tubes and inlet tube of the pump matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo enough to cause a separation.A separation of this type could then allow the loss of fluid, making the device inoperable.
|
| |
|