| Model Number |
GI100 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The united kingdom pathologist reported the event of "no staining for b-catenin".The test section was found negative at the initial hospital.This case was then sent to another hospital that carried out b-catenin and the test section showed positive nuclear staining.Controls stained correctly.Diagnostics were not altered.No harm or patient impact was indicated.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is currently ongoing.No additional information has been made available.Therefore, this report is being filed as part of agilent's commitment to due diligence reporting.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the log files did not show any instrument anomalies which could have caused the reported issue.It has not been possible to gather any additional information from the customer to support further investigation.Root cause of this problem could not be identified.
|
| |
|
Additional Manufacturer Narrative
|
|
G3: date received by manufacturer was updated with correct information.
|
| |
|