• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. DAKO OMNIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. DAKO OMNIS Back to Search Results
Model Number GI100
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
The united kingdom pathologist reported the event of "no staining for b-catenin".The test section was found negative at the initial hospital.This case was then sent to another hospital that carried out b-catenin and the test section showed positive nuclear staining.Controls stained correctly.Diagnostics were not altered.No harm or patient impact was indicated.
 
Additional Manufacturer Narrative
Investigation is currently ongoing.No additional information has been made available.Therefore, this report is being filed as part of agilent's commitment to due diligence reporting.
 
Event or Problem Description
Additional information was received that the log files did not show any instrument anomalies which could have caused the reported issue.It has not been possible to gather any additional information from the customer to support further investigation.Root cause of this problem could not be identified.
 
Additional Manufacturer Narrative
G3: date received by manufacturer was updated with correct information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAKO OMNIS
Common Device Name
DAKO OMNIS
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, 76892 3
SN  768923
Manufacturer (Section G)
SHANDON DIAGNOSTICS LIMITED
tudor road
manor park
runcorn, WA7 1 TA
UK   WA7 1TA
Manufacturer Contact
mary o'neill
1834 state highway 71 west
cedar creek, TX 78612
3026338510
MDR Report Key16626140
Report Number3003423869-2023-00045
Device Sequence Number4551204
Product Code KPA
UDI-Device Identifier05700573004904
UDI-Public05700573004904
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGI100
Device Catalogue NumberGI10030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2023
Supplement Date Received by Manufacturer02/27/2023
04/12/2023
Initial Report FDA Received Date03/28/2023
Supplement Report FDA Received Date04/24/2023
10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-