MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INBONE STEM, TIBIAL, TOP SIZE, 16 MM, RIGHT & LEFT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INBONE STEM, TIBIAL, TOP SIZE, 16 MM, RIGHT & LEFT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 200011902
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Health Effect - Clinical Codes Unspecified Infection (1930); Inadequate Osseointegration (2646)
Date of Event 03/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a total ankle replacement.Allegedly, the patient underwent an explant surgery due to infection.The patient currently has a bone cement spacer.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.Device disposition unknown.
 
Event or Problem Description
It was reported that the patient underwent a total ankle replacement.Allegedly, the patient underwent an explant surgery due to infection.The patient currently has a bone cement spacer.
 
Additional Manufacturer Narrative
The reported event could be confirmed based on available medical record and health care expert¿s opinion.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed- ¿as there is a girdlestone situation without a tar visible the assessment is limited.There is a large bone loss visible in the situation.¿ based on investigation, the root cause was attributed to a patient related issue.The failure was caused by loss of huge portion of bone resulting in a girdlestone situation.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE STEM, TIBIAL, TOP SIZE, 16 MM, RIGHT & LEFT
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16627039
Report Number3010667733-2023-00150
Device Sequence Number4426624
Product Code HSN
UDI-Device Identifier00840420123909
UDI-Public00840420123909
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K051023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number200011902
Device Catalogue Number200011902
Device Lot Number1670258
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2023
Supplement Date Received by Manufacturer07/05/2023
Initial Report FDA Received Date03/28/2023
Supplement Report FDA Received Date07/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
-
-