| Model Number |
60-05-00 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Livanova deutschland manufactures the s5 erc tubing clamp / 500mm.The incident occurred in zürich, switzerland.In order to fix the issue, photo sensor was replaced at manufacturer's site.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova deutschland received a report that a s5 erc (electrical remote-controlled) tubing clamp / 500mm could no longer be opened during the operation.The affected part was returned to manufacturer's site for a detailed investigation.The investigator could reproduce the reported issue.During test, the unit gave an error message related to motor jammed.There was no patient injury.The clamp.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaints database analysis revealed no similar event since unit installation in 2017.Based on the above facts and considering repair activity results, the root cause of the reported event was traced back to defective photo sensors.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|