• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 60MM 44CM SHAFT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 60MM 44CM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Model Number PLE60A
Medical Device Problem Codes Failure to Form Staple (2579); Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch#: unk.A manufacturing record evaluation was performed for the finished device lot/batch number: u95560, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the surgery.After the fifth fired, the surgeon noted that the tissue was not cut and staples deployed on the tissue, noted the staples malformed.Changed another device to complete the procedure, there was no patient consequence reported.No additional information could be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 4/19/2023.D4: batch # unk.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ple60a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The cut line was completed and the staples meet the staple form release criteria.However, the cut was noted to be jagged due to a damaged knife.One possible cause for this type of damage to the knife is when the device is fired over an already existing staple line, hard object, or thicker tissue than indicated.Repeatedly firing across existing staple lines can also reduce the ability to cut cleanly.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number u5dv3w, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD ECH FLEX 60MM 44CM SHAFT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key16630452
Report Number3005075853-2023-02017
Device Sequence Number17473815
Product Code GDW
UDI-Device Identifier10705036014300
UDI-Public10705036014300
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberPLE60A
Device Catalogue NumberPLE60A
Device Lot NumberU95560
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/03/2023
Supplement Date Received by Manufacturer03/29/2023
Initial Report FDA Received Date03/28/2023
Supplement Report FDA Received Date04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1