Model Number 100014514
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2023
Event Type  malfunction  
Manufacturer Narrative
One ensite velocity¿ amplifier was received for evaluation.Power was applied to the amplifier and the amplifier did not initiate the power-on-self-test (post).The amplifier stayed status steady amber ¿not-ready¿ and could not communicate with the test ensite computer.Internal inspection revealed the ic chip on the ampro assembly (part of the micro-processor board) was missing due to an over temperature condition, which duplicated the reported symptom.When the micro-processor board was replaced with a known good processor, the amplifier passed the post and communicated with the test station.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.The reported event was able to be duplicated by power sequencing and internal inspection.Based on the information provided to abbott and the investigation performed, the root cause was attributed to a thermal event on the ampro assembly of the micro-processor board.
 
Event Description
This report is to advise of a thermal event noted during analysis,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16630627
MDR Text Key312159122
Report Number2184149-2023-00071
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100014514
Device Catalogue Number100014514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/23/2023
Initial Date FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-