ALPHATEC SPINE, INC. INVICTUS; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 15054-065-040;15100
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Around 4-weeks postoperatively, a set screw had become disengaged.A revision surgery was performed to remove the hardware.It was reported the thread of the screw was damaged.
 
Additional Manufacturer Narrative
The returned implants are currently being evaluated.A follow-up report with the results of the investigation will be submitted upon completion.
 
Additional Manufacturer Narrative
H3.Pn 15054-05-040 ln ap0036: 1/18/2022 pn 15100 ln ap00820: 08/12/2022 h6.Investigation findings: 3243 investigation conclusions: 61 h10.Review of the device history record for the fixed screw, 06.5 x 40mm (pn 15054-065-04 ln ap00436) indicates the implants conformed to dimensional specifications at the time of manufacturing and passed inspection requirements with no non-conformities reported.There were no issues during the manufacture of this product that would contribute to this complaint condition.Review of the device history record for qty 3 set screws (pn 15100 ln ap00820) indicates the implants conformed to dimensional specifications at the time of manufacturing and passed inspection requirements with no non-conformities reported.There were no issues during the manufacture of this product that would contribute to this complaint condition.The fixed screw pn 15054-065-040 was returned with wear patterns observed on the rod slot, external body of the tulip, and the proximal helical flange thread.The proximal helical flange is damaged and there is wear on the single side of rod slot.The three set screws pn 15100 returned with displayed wear on the underside (rod-interfacing side) face as well as a deformed dimple feature.The sweeping wear patterns are a result of the set screw smearing against the implanted rod during set screw driving and final tightening.Root cause and failure mode: non-normal rod-tulip interface the most likely cause of the set screw backout and eventual dislocation is due to a non-normal rod to tulip interface caused by insufficient rod contouring at the tulip interface.The rod slot face on the returned fixed screw displays wear on only one side, indicative of single sided engagement between the rod and screw (i.E.Non-normal interface).The wear patterns of set screw 1 also support an asymmetric loading where the left side of the set screw will show wear before the right side of the set screw.The expected wear pattern would be a significantly longer streak of wear on one side of the set screw relative to the other as one side of the set screw drags much more than the contralateral side.Non-normal rod-tulip engagement with set screw: because there exists a gap between the rod and tulip, the rod is able to have micromotions causing the set screw to begin backing out.Eventually, the set screw will become unthreaded until there exists only a partial thread engagement.At this point, the tulip thread can be sheared or peeled up due to extreme loading under insufficient thread engagement.Once the thread shears, the set screw can become completely disassociated form the tulip.The implants were used for treatment, not diagnosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVICTUS
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vide roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC
1950 camino vide roble
carlsbad CA 92008
Manufacturer Contact
wesley channell
1950 camino vide roble
carlsbad, CA 92008
9014283693
MDR Report Key16632126
Report Number2027467-2023-00025
Device Sequence Number20228178
Product Code KWP
UDI-Device Identifier00190376223582
UDI-Public(01)00190376223582(10)AP00436;(01)00190376137902(10)AP00820
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K181677
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15054-065-040;15100
Device Catalogue Number15054-065-040;15100
Device Lot NumberAP00436;AP00820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2023
Supplement Date Received by Manufacturer02/28/2023
Initial Report FDA Received Date03/28/2023
Supplement Report FDA Received Date08/21/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-