| Model Number |
TG5412545 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted in the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil (subject device) was fully implanted inside the aneurysm.During a hand injection it was noticed that the helical portion came down into the parent vessel and then retracted back into the anx within the same run.The procedure was completed successfully with the same coil (subject device).No clinical consequences were reported to the patient due to this event.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil (subject device) was fully implanted inside the aneurysm.During a hand injection it was noticed that the helical portion came down into the parent vessel and then retracted back into the anx within the same run.The procedure was completed successfully with the same coil (subject device).No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint.
|
| |
|