• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number FINETUNER ECHO [FINETUNER 2]
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
 
Event or Problem Description
A finetuner echo was returned to service and repair due to a defect.
 
Additional Manufacturer Narrative
Conclusion: according the information received, a finetuner echo was sent to service _ repair due to not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
 
Event or Problem Description
A finetuner echo was returned to service and repair due to a defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key16633665
Report Number9710014-2023-00257
Device Sequence Number20362728
Product Code MCM
UDI-Device Identifier09008737344290
UDI-Public(01)09008737344290
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberFINETUNER ECHO [FINETUNER 2]
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2023
Supplement Date Received by Manufacturer03/27/2023
Initial Report FDA Received Date03/29/2023
Supplement Report FDA Received Date04/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
-
-