| Model Number |
G5S-01A |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device's defib output was out of specification.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report was observed during review of the device logs.However, the device was put through extensive testing including bench handling, power cycling, and functional stress testing without duplicating the report.The error was cleared and did not reoccur.The main board and hv capacitors were replaced as a precaution.The device was recertified and returned to the customer.The electrode pads used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
|
| |
|