| Model Number |
8881513231 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported it looks like the rubber plungers melted or something.When they tried to pull up a vaccine the contents dripped out of the plunger side of the syringe.They kind of looked flipped to the side.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information available to us, we were able to confirm the event, and determined that: the device history record was reviewed and indicated that the product was released accomplishing all quality standards.Five unused samples were returned for evaluation and the reported issue was seen during visual evaluation.A review of all available information was unable to determine an exact root cause.No corrective actions required at this time.All information received will be used for further tracking and trending purposes.
|
| |
|