BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 87047
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that during an ablation procedure to treat atrial fibrillation (afib), an intellanav mifi oi catheter was selected for use.During ablation the catheter could not be visualized on the rhythmia hdx mapping system.The ablation cable was replaced, and this did not resolve the problem.The ablation catheter was replaced and the problem was resolved.Upon removal and inspection of the catheter a breach in the catheter shaft was found.During the procedure no resistance was felt while maneuvering this catheter.The damage was not noticed before the catheter was inserted into the patient.No fragments were left in the patient and no interventions were needed to retrieve any pieces of the device from inside the body.The catheter was replaced and the procedure was completed without any patient complications.The catheter is expected to be returned for laboratory analysis.
 
Additional Manufacturer Narrative
Additional information received on 28mar2023, indicated that there was an intermittent magnetic tracking failure.The mapping system used was rhythmia hdx, software version: 4.5.No other error messages were observed during the procedure.This intellanav mifi oi catheter was received by boston scientific and a laboratory analysis was performed.Visual inspection revealed a cut in the catheter shaft.Functional testing was performed; the steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Electrical and continuity testing were performed, and the device showed all results were within specifications.Additional testing was conducted, the device was able to complete an ablation without any difficulties.Laboratory analysis was able to confirm the reported clinical observation regarding "tears/rips/holes/sep dev matrl" was confirmed.For the "tracking/accuracy issue" event the device showed no evidence or defects that could have contributed to this specific issue.
 
Event or Problem Description
It was reported that during an ablation procedure to treat atrial fibrillation (afib), an intellanav mifi oi catheter was selected for use.During ablation the catheter could not be visualized on the rhythmia hdx mapping system.The ablation cable was replaced, and this did not resolve the problem.The ablation catheter was replaced and the problem was resolved.Upon removal and inspection of the catheter a breach in the catheter shaft was found.During the procedure no resistance felt while maneuvering this catheter.The damage was not noticed before the catheter was inserted into the patient.No fragments were left in the patient and no interventions were needed to retrieve any pieces of the device from inside the body.The catheter was replaced and the procedure was completed without any patient complications.The catheter has been returned for laboratory analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV MIFI OPEN-IRRIGATED
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key16637752
Report Number2124215-2023-13867
Device Sequence Number12067146
Product Code OAD
UDI-Device Identifier08714729938385
UDI-Public08714729938385
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P150005/S017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2024
Device Model Number87047
Device Catalogue Number87047
Device Lot Number0030557241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/03/2023
Supplement Date Received by Manufacturer03/30/2023
Initial Report FDA Received Date03/29/2023
Supplement Report FDA Received Date04/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-