Model Number 5052
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that as an employee was unloading their amsco 5052 washer/disinfector, the door of the unit closed on the employee's hand resulting in an injury.Medical treatment was sought and administered, and the employee returned to work.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the amsco 5052 washer/disinfector and found it to be operating according to specification.The technician stated that the door was functioning normally and closing at the correct speed.The technician also found that there were no issues with the function of the door safety switches.No repairs were required, and the unit was returned to service.The reported event is attributed to user error as the employee should have ensured that their hand was away from the unit's threshold when the door of the unit was closing.A steris account manager conducted in-service training with user facility personnel on the proper use and operation of the amsco 5052 washer/disinfector, specifically on safe operating protocols.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO 5052 WASHER/DISINFECTOR
Common Device Name
WASHER/DISINFECTOR
Manufacturer (Section D)
STERIS CANADA ULC
490 boulevard armand-paris
quebec, G1C 8 A3
CA  G1C 8A3
Manufacturer (Section G)
STERIS CANADA ULC
490 boulevard armand-paris
quebec, G1C 8 A3
CA   G1C 8A3
Manufacturer Contact
daniel davy
5960 heisley rd
mentor, OH 44060
4403927453
MDR Report Key16638068
Report Number9680353-2023-00008
Device Sequence Number12121319
Product Code MEC
UDI-Device Identifier00724995192976
UDI-Public00724995192976
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5052
Device Catalogue NumberFH14072
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2023
Initial Report FDA Received Date03/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-