| Model Number |
324912 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that 8 bd veo¿ insulin syringes with the bd ultra-fine¿ needle had mixed products including in their box.This complaint was created to capture the 1st of 2 related incidents.The following information was provided by the initial reporter: " mixed components, needle bends during use, bent prior, and injection causes bump.The original complaint reported by the consumer that there are mixed components in package, needle bends during injection, some pen needles are bent prior to injection.Injection causes a "bump under his skin."".
|
| |
|
Additional Manufacturer Narrative
|
|
H6: unable to perform complaint lot history check for leakage due to unknown lot number.The occurrence is unknown since the lot number is unknown.This will be considered the 1st related complaint for guidance on the investigation.Investigation summary: no samples (including photos) were returned for the unknown lot#, therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Due to the batch being unknown, no dhr review can be completed.As no samples and/or photo(s) were received the investigation concluded: - unconfirmed: embecta was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned root cause cannot be determined at this time as the issue is unconfirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported that 8 bd veo¿ insulin syringes with the bd ultra-fine¿ needle had mixed products including in their box.This complaint was created to capture the 1st of 2 related incidents.The following information was provided by the initial reporter: " -mixed components.-needle bends during use.-bent prior.-injection causes bump.The original complaint reported by the consumer that there are mixed components in package, needle bends during injection, some pen needles are bent prior to injection.Injection causes a "bump under his skin.".
|
| |
|