• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH BLUNT TIP SCREW, 4X46MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH BLUNT TIP SCREW, 4X46MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number N/A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/08/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant products: proximal humerus, right, 11x160mm; item#47249616011; lot#33082002.Blunt tip screw, 4x36mm; item#47248603640; lot#3091452.Blunt tip screw, 4x48mm; item#47248604840; lot#3091539.Torque limiting handle, a.5; item#27923; lot#unknown.Report source: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2023 - 00137, 0009613350 - 2023 - 00138, and 0009613350 - 2023 - 00140.
 
Event or Problem Description
It was reported that 4 weeks post-implantation, it was confirmed by x-rays that the lag screw was sliding to outer side.The surgeon will continue to monitor the patient's condition.Revision surgery is not currently planned.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.An undated radiograph was received and assessed by a radiologist.The single ap view of the right shoulder shows an intramedullary nail and screw fixation traverses a proximal humeral fracture with anatomic alignment.Two of the more proximal screws are malpositioned with the screw heads well proximal to the humeral cortex.However, as no other radiographs are provided, a comparison with an earlier time point is not possible.Fracture lucency is noted.Bone quality is markedly osteopenic.Different factors, related to either patient condition, behavior or implantation procedure, might have led or contributed to the migration of the screws.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus natural nail proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, 4X46MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16641950
Report Number0009613350-2023-00139
Device Sequence Number14790560
Product Code HSB
UDI-Device Identifier00889024505476
UDI-Public(01)00889024505476(17)260923(17)3082116
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827 / K2
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/23/2026
Device Model NumberN/A
Device Catalogue Number47-2486-046-40
Device Lot Number3082116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2023
Supplement Date Received by Manufacturer04/28/2023
Initial Report FDA Received Date03/30/2023
Supplement Report FDA Received Date05/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
-
-