ZIMMER SWITZERLAND MANUFACTURING GMBH PROXIMAL HUMERUS, RIGHT, 11X160MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products:¿ blunt tip screw, 4x36mm; item#47248603640; lot#3091452, blunt tip screw, 4x46mm; item#47248604640; lot#3082116, blunt tip screw, 4x48mm; item#47248604840; lot#3091539, torque limiting handle, a.5; item#27923; lot#unknown.Japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2023 - 00138, 0009613350 - 2023 - 00139, 0009613350 - 2023 - 00140.
|
| |
|
Event or Problem Description
|
|
It was reported that 4 weeks post-implantation, it was confirmed by x-rays that the lag screw was sliding to outer side.The surgeon will continue to monitor the patient's condition.Revision surgery is not currently planned.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.An undated radiograph was received and assessed by a radiologist.The single ap view of the right shoulder shows an intramedullary nail and screw fixation traverses a proximal humeral fracture with anatomic alignment.Two of the more proximal screws are malpositioned with the screw heads well proximal to the humeral cortex.However, as no other radiographs are provided, a comparison with an earlier time point is not possible.Fracture lucency is noted.Bone quality is markedly osteopenic.Different factors, related to either patient condition, behavior or implantation procedure, might have led or contributed to the migration of the screws.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus natural nail proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|