• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER Back to Search Results
Model Number 119110
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the foley catheter spontaneously fell out of the patient due to water leak.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter spontaneously fell out of the patient due to water leak.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The reported failure is within specification as the reported failure could not be reproduced.The product was used for patient treatment or diagnosis.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.Visual evaluation of the returned sample noted one opened (without original packaging), a 3-way temperature sensing catheter.Visual inspection noted no defects on the catheter.The catheter balloon was inflated with 3.5 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated with no cuffing or leaks noted.This is within specification per inspection procedure ip7603444, revision 0, which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.As the reported event is unconfirmed a labeling review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16642856
Report Number1018233-2023-02150
Device Sequence Number20181668
Product Code EZL
UDI-Device Identifier00801741046001
UDI-Public(01)00801741046001
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number119110
Device Catalogue Number119110
Device Lot NumberNGGV5074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/08/2023
Supplement Date Received by Manufacturer05/05/2023
Initial Report FDA Received Date03/30/2023
Supplement Report FDA Received Date05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-