| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cmp-(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-00672, and 0001825034-2023-00673.Report source: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee procedure, as the packaging to the femoral pegs was opened, the inner sterile bag tore, and the implants dropped.An issue was noted with the sterile packaging on two additional implants.The procedure was completed using a fourth set of pegs.No adverse events have been reported as a result of the malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h10.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event cannot be determined.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned product found the inner sterile pouch is broken into multiple pieces.Evaluation of the outer pouch found no damage.Sterility has not been breached.The outer carton exhibits crush damage; however this failure was not initially reported and likely occurred during transit back to the manufacturer.Device history record was reviewed and no discrepancies were found.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event can be attributed to the supplied sterile inner pouch that has become brittle due to length of time in the field.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|